Origin: Korea.
Performance
- Microprocessor controller and temperature calibration function allows for high precision temperature control.
- Sterilization temperature at 110 to 123°C
- Melting* temperature at 60 to 100°C
* Liquefaction of coagulated agar media. - Automatic sterilization system for unattended operation.
- Heating > Check safe > Sterilization > Sterilization safe > Exhausting > End - Convenient operation modes for various uses.
- Standard modes and program modes are listed in detail as below tables. - Temperature calibration function.
Standard modes | Temp. (°C) | Time (min.) |
---|---|---|
Decontamination | 121 | 35 |
Solids sterilization | 121 | 20 |
Liquids sterilization | 121 | 15 |
Melting | 100 | 30 |
Program modes | Temp. (°C) | Time (min.) |
---|---|---|
Sterilization 1 | 110 to 123 | 1 to 999 |
Sterilization 2 | 110 to 123 | 1 to 999 |
Liquid sterilization | 110 to 123 | 1 to 999 |
Melting | 60 to 100 | 1 to 999 |
Convenience
- Conveniently monitored by an independent control display, sterilization process display, and pressure gauge.
- VFD for visualization of temp, pressure (optional), timer, warning alert checks.
- LED lamps for sterilization process check.
- Easy pressure check with the standard gauge attached to the front. - Vertical type autoclaves which allow for top loading are suitable for high volume sterilization.
- Baskets are stackable two high or even more levels, which secure inside the chamber space requirements. - Low-maintenance
- The chamber is constructed of stainless steel which is superiorly resistant to corrosion.
- Easy to replace silicone gasket. - Convenient documentation with the optional data recorder.
- Dot type recorder(6 point) allows for precise documentation.
Safety
- Mechanical safety devices
- Over current protection.
- Over temperature protection by automatic power cutoff.
- Over pressure protection by automatic and manual safety valve. - Safety warning systems
- Over temperature warning.
- Low temperature warning.
- Low water level warning.
- Low heat warning if the temperature does not reach the sterilization or melting temperature.
- Sterilization fail warning if the operation is unexpectedly stopped during sterilization process. - Slide opening door with plastic cover of low thermal conductivity prevents burns from the contact.